CAS Number | 1491-59-4 |
---|---|
Molecular Formula | C16H24N2O |
Molecular Weight | 260.382 |
InChI Key | WYWIFABBXFUGLM-UHFFFAOYSA-N |
LogP | 4.01 |
Synonyms |
|
Applications:
HPLC Separation of EDTA, Oxymetazoline and Benzyl Alcohol in Nasal Spray
September 4, 2020
HPLC Method for EDTA (Ethylenediaminetetraacetic Acid), Oxymetazoline, Benzyl alcohol, Oxymetazoline hydrochloride on Newcrom BH by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of EDTA, Oxymetazoline and Benzyl Alcohol
Benzyl alcohol is found in over-the-counter medications, topical creams, lotions, shampoos, and facial cleansers as an antibacterial, preservative, and/or fungicide. Benzyl alcohol is a colorless liquid with a mild pleasant aromatic scent. Benzyl alcohol is found in many essential oils including jasmine, hyacinth, neroli, rose, and ylang-ylang. Oxymetazoline is a decongestant that shrinks blood vessels in the nasal passages. All active compounds of nasal spray can be separated in HPLC on a Newcrom BH mixed-mode column. The analytical method uses acetonitrile (ACN) gradient and water with sulfuric acid (H2SO4) as buffer and UV detected at 260nm.
Column | Newcrom BH, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | Gradient MeCN – 2- 70%, 10 min |
Buffer | H2SO4 – 0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV 260nm |
Class of Compounds | Acid, Hydrophilic, Preservatives |
Analyzing Compounds | EDTA (Ethylenediaminetetraacetic Acid), Oxymetazoline, Benzyl alcohol, Oxymetazoline hydrochloride |
Application Column
Newcrom BH
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
EDTA (Ethylenediaminetetraacetic Acid)
Oxymetazoline
Oxymetazoline hydrochloride
Improved Peak Shape of a Basic Compound on Primesep B2 Column
June 10, 2004
Primesep B2 shows superior efficiency, symmetry, and peak shape for a hydrophilic basic compound compared to two leading reversed-phase C18 columns. Primesep B2 contains an embedded basic functional group that shields the underlying silica by ionic repulsion with the analyte. On reversed-phase columns exposed active silanols provide unintended polar interactions with basic analytes that lead to tailing. The HPLC separation uses a simple mobile phase of water, acetonitrile (MeCN, ACN) and trifluoracetic acid (TFA) with UV detection at 250 nm.
Column | Primesep B2, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O |
Buffer | TFA |
Flow Rate | 1.0 ml/min |
Detection | UV, 250 nm |
Class of Compounds |
Drug, Acid, Hydrophilic, Ionizable, Vitamin, Supplements, Amino acid |
Analyzing Compounds | Oxymetazoline |
Application Column
Primesep B2
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.