CAS Number | 51-48-9 |
---|---|
Molecular Weight | 776.87 |
Synonyms |
|
Applications:
UV-Vis Spectrum of Thyroxine
July 23, 2024
For optimal results in HPLC analysis, it is recommended to measure absorbance at a wavelength that matches the absorption maximum of the compound(s) being analyzed. The UV spectrum shown can assist in selecting an appropriate wavelength for your analysis. Please note that certain mobile phases and buffers may block wavelengths below 230 nm, rendering absorbance measurement at these wavelengths ineffective. If detection below 230 nm is required, it is recommended to use acetonitrile and water as low UV-transparent mobile phases, with phosphoric acid and its salts, sulfuric acid, and TFA as buffers.
For some compounds, the UV-Vis Spectrum is affected by the pH of the mobile phase. The spectra presented here are measured with an acidic mobile phase that has a pH of 3 or lower.
HPLC Method for Analysis of Iodide and Thyroid Hormones
October 30, 2018
Iodine is an important element in production of thyroid hormones, T3 and T4, which are essential regulators of organism’s metabolism. The glands’ concentration of Iodide, hormones T3 and T4, along with their precursors: 3,5-Diiodo-L-thyronine (T2), 3,5-Diiodo-L-tyrosine (DIT), and 3-Iodo-L-tyrosine (MIT) can now be studied simultaneously with this HPLC method. In order to analyze the hydrophobic hormones along with iodide (a very polar inorganic ion) a special mixed-mode column, Primesep D, was used. This column provides both strong hydrophobic and anion exchange properties. The Primesep D silica-bonded ligand is comprised of a long alkyl chain and an embedded amino functional group. The TFA was used as an ionic modifier to provide stable acidic pH and sufficient ion strength for the mobile phase. The mobile phase composition is suitable for UV, MS, ELSD, and CAD detection.
Column | Primesep D, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN |
Buffer | TFA |
Flow Rate | 1.0 ml/min |
Detection | UV, 230 nm |
Class of Compounds |
Drug, Hydrophilic, Ionizable, Supplements |
Analyzing Compounds | T3, T4, 3,5-Diiodo-L-thyronine (T2), 3,5-Diiodo-L-tyrosine (DIT), and 3-Iodo-L-tyrosine (MIT) |
Application Column
Primesep D
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select options3,5-Diiodo-L-tyrosine dihydrate (DIT)
3-Iodo-L-tyrosine (MIT)
Iodide
Iodine
Levothyroxine (T4)
Liothyronine (T3)
Methimazole
HPLC Method for Analysis of Levothyroxine Sodium Tablets
June 11, 2018
Column | Primesep D, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 30/70% |
Buffer | TFA – 0.1 % |
Flow Rate | 1.0 ml/min |
Detection | UV, 230 nm |
Class of Compounds |
Drug, Hormone, Acid, Hydrophobic |
Analyzing Compounds | Levothyroxine |
Application Column
Primesep D
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select options