Separation type: Liquid Chromatography Reverse Phase.
HPLC Method for Analysis of Cannabidiol (CBD), Cannabinol (CBN), Tetrahydrocannabinol (THC), Delta-8-Tetrahydrocannabinol (8-THC), Delta-9-Tetrahydrocannabinol (THC), Tetrahydrocannabinolic acid (THCA-A) on Cannsep C by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of CBD, CBN, THC, 8-THC and THCA.
Δ8-THC, Δ9-THC, CBD, THCA and CBN are cannabinoids found in Cannabis plants. They have some medical uses among which is the prevention of nausea and vomiting caused by some cancer medications. They can be separated using HPLC on a reverse-phase Cannsep C column using a mobile phase consisting of methanol (MeOH), Acetonitrile (MeCN, ACN), and water with sulfuric acid (H2SO4) as buffer and UV detected at 210nm.
Column | Cannsep C, 2.1 x 100 mm, 3 µm, 100 A |
Mobile Phase | MeCN/MeOH/H2O – 45/40/15% |
Buffer | H2SO4 – 0.2% |
Flow Rate | 0.2 ml/min |
Detection | UV, 200 nm |
Class of Compounds | Drug, Hydrophilic, Ionizable, Supplements |
Analyzing Compounds | Cannabidiol (CBD), Delta-9-Tetrahydrocannabinol (THC), CBD, CBN, Δ9-THC, Δ8-THC, Tetrahydrocannabinolic acid (THCA) |
Application Column
Cannsep C
Column Diameter: 2.1 mm
Column Length: 100 mm
Particle Size: 3 µm
Pore Size: 100 A
Cannabinol (CBN)
Delta-8-Tetrahydrocannabinol (8-THC)
Delta-9-Tetrahydrocannabinol (THC)
Tetrahydrocannabinol (THC)
Tetrahydrocannabinolic acid (THCA-A)