CAS Number | 57-00-1 |
---|---|
Molecular Formula | C4H9N3O2 |
Molecular Weight | 131.135 |
InChI Key | CVSVTCORWBXHQV-UHFFFAOYSA-N |
LogP | -2.16 |
Synonyms |
|
Applications:
HPLC Method for Analysis of Creatine and Creatinine on Primesep 200 Column
December 10, 2024
HPLC Method for Analysis of Creatine, Creatinine on Primesep 200 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Creatine, Creatinine
Creatine is a naturally occurring compound found in muscles and is used to produce energy during high-intensity exercise. It is synthesized in the body from amino acids and stored in muscles for quick energy release. Creatinine, on the other hand, is a waste product produced from the breakdown of creatine. It is filtered by the kidneys and excreted in urine. The measurement of creatinine levels in the blood and urine is commonly used to assess kidney function, as high levels may indicate impaired kidney function or other health issues.
Creatine and Creatinine can be analyzed and separated using a Primesep 200 mixed-mode stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water, acetonitrile (MeCN), and phosphoric acid as a buffer. Detection is carried out using UV.
Column | Primesep 200, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 10/90% |
Buffer | H3PO4 – 0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV 210 nm |
Sample | 1. 0.2 mg/mL, 2. 0.3 mg/mL |
Injection volume | 5 µl |
LOD* | 1. 46.5 ppb, 2. 27.4 ppb |
Class of Compounds | Quaternity amines |
Analyzing Compounds | Creatine, Creatinine |
Application Column
Primesep 200
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Creatinine
HPLC Separation of Creatine, Creatinine and Histidine with MS Conditions
January 13, 2005
Creatine, creatinine, and histidine are all important amino acids in the human body. All of these can be separated by liquid chromatography. The mobile phase contains acetonitrile (MeCN), water, and phormic acid. Smaller 3 µm particle columns are available for fast UPLC applications. This liquid chromatography method is scalable and can be used for isolation of impurities in preparative separation. It also suitable for pharmacokinetics.
Column | Primesep 200, 3.0×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 20/80% |
Buffer | Formic Acid – 0.1% |
Flow Rate | 0.5 ml/min |
Detection | UV, 210 nm |
Class of Compounds |
Drug, Acid, Hydrophilic, Ionizable, Hormone |
Analyzing Compounds | Creatine, Creatinine, Histidine |
Application Column
Primesep 200
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select optionsCreatinine
Histidine
HPLC Separation of Creatine, Creatinine and Histidine
January 13, 2005
Creatine is produced by the body and plays a major role in recycling ATP. It also stabilizes pH in certain tissues. Creatine can be analyzed by this reverse phase (RP) HPLC method with simple conditions. The mobile phase contains acetonitrile (MeCN), water, and TFA. For Mass-Spec (MS) compatible applications the phosphoric acid needs to be replaced with formic acid. Smaller 3 µm particle columns are available for fast UPLC applications. This liquid chromatography method is scalable and can be used for isolation of impurities in preparative separation. It also suitable for pharmacokinetics.
Column | Primesep 200, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 1/90% |
Buffer | TFA -0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV, 210 nm |
Class of Compounds |
Drug, Acid, Hydrophilic, Ionizable, Hormone |
Analyzing Compounds | Creatine, Creatinine, Histidine |
Application Column
Primesep 200
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.