CAS Number | 55-18-5 |
---|---|
Molecular Formula | C4H10N2O |
Molecular Weight | 102.137 |
InChI Key | WBNQDOYYEUMPFS-UHFFFAOYSA-N |
LogP | 0.5 |
Synonyms |
|
Applications:
HPLC Method for Detection of NDEA in Valsartan Drug Substance
November 14, 2018
HPLC Method for Valsartan, N-Nitrosodiethylamine on Newcrom R1 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Valsartan, N-Nitrosodiethylamine
Valsartan products are used to treat high blood pressure and congestive heart failure. On July 13, 2018, FDA announced a recall of valsartan tablets because of the potential for certain products to contain an impurity, N-nitrosodimethylamine (NDMA). Nitrosamine impurities is classified as a probable human carcinogen and is believed to have been introduced into the finished products as a result of the manufacturing process of the drug substance.
SIELC developed simple and fast HPLC method to measure presence of NDEA in drug formulations. This UV based method allows to detect ppb level of the compound in solution. Simple mobile phase comprised of acetonitrile (MeCN) and water with detection at 232 nm using Newcrom R1 column can be used to detect both the API and the NDEA.
Column | Newcrom R1, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | Gradient MeCN – 10-75%, 12 min |
Buffer | H3PO4- 0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV, 232 nm |
Class of Compounds |
Drug, Carcinogen, Zwitterionic, Hydrophilic, Hydrophobic, Ionizable |
Analyzing Compounds | Valsartan, N-Nitrosodiethylamine |
Application Column
Newcrom R1
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
Valsartan
HPLC Method for Analysis of N-Nitrosodiethylamine on Newcrom R1 Column
November 14, 2018
HPLC Method for N-Nitrosodiethylamine on Newcrom R1 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of N-Nitrosodiethylamine
N-Nitrosodiethylamine (NDEA) or Diethylnitrosamine is potential carcenogenic agent was found in some blood pressure medications. Those medications were recalled by FDA with the concern of high concentration of the chemical.
SIELC developed simple and fast HPLC method to measure presence of NDEA in drug formulations. This UV based method allows to detect ppb level of the compound in solution. Simple mobile phase comprised of acetonitrile (MeCN) and water with detection at 2** nm using Newcrom R1 column can be used to detect both the API and the NDEA.
Column | Newcrom R1, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | MeCN – 20% |
Buffer | No |
Flow Rate | 1.0 ml/min |
Detection | UV, 232 nm |
Class of Compounds | Carcinogen, Zwitterionic, Hydrophobic |
Analyzing Compounds | N-Nitrosodiethylamine |
Application Column
Newcrom R1
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended