HPLC Method for Analysis of Econazole on Primesep 200 by SIELC Technologies

Econazole is an antifungal medication with the chemical formula C18H15Cl3N2O. It is primarily used to treat a variety of fungal skin infections including but not limited to athlete’s foot, tinea, ringworm, and jock itch. It belongs to the imidazole class of antifungal agents. Like other imidazole antifungals, econazole works by inhibiting the enzyme responsible for producing ergosterol, a key component of the fungal cell membrane. A lack of ergosterol disrupts the integrity of the fungal cell membrane, leading to the death of the fungus. You can find detailed UV spectra of Econazole and information about its various lambda maxima by visiting the following link.
Econazole can be retained, and analyzed on a Primesep 200 mixed-mode stationary phase column using an isocratic analytical method with a simple mobile phase of water, Acetonitrile (MeCN), and a sulfuric acid as a buffer. This analysis method can be detected in the UV 200 nm.
| Column | Primesep 200, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | Gradient MeCN – 40-70%, 10 min |
| Buffer | H2SO4 – 0.2% |
| Flow Rate | 1.0 ml/min |
| Detection | UV, 200 nm |
| Peak Retention Time | 6.89 min |
| Class of Compounds | Allylamines |
| Analyzing Compounds | Econazole |
Application Column
Primesep 200
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended