HPLC Method for Acetaminophen (Paracetamol) on Primesep 100 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Acetaminophen (Paracetamol)
Acetaminophen, also known as Paracetamol, is a medication with the molecular formula C8H9NO2. It is an over-the-counter pain killer that also reduces fever. Unlike Ibuprofen and naproxen, it is not an anti-inflammatory drug.
Acetaminophen (Paracetamol) can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.
Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | MeCN – 10% |
Buffer | H2SO4 – 0.2% |
Flow Rate | 1.0 ml/min |
Detection | UV 275 nm |
Class of Compounds | Organic Compound |
Analyzing Compounds | Acetaminophen (Paracetamol) |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended