High Performance Liquid Chromatography (HPLC) Method for Analysis of Sildenafil or Viagra on Primesep 100 by SIELC Technologies.
Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of Sildenafil or Viagra
Sildenafil Citrate (also known as Viagra) is one of the most popular medications used to treat erectile dysfunction and pulmonary hypertension. Sildenafil can be retained and analyzed on a Primesep 100 mixed-mode stationary phase column using an isocratic analytical method with a simple mobile phase of water, Acetonitrile (MeCN), and a Sulfuric acid (H2SO4) buffer. This analysis method can be detected in the UV regime at 214 nm and 294 nm with high resolution and peak symmetry.
Condition
Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN – 60% |
Buffer | H2SO4 – 0.2% |
Flow Rate | 1.0 ml/min |
Detection | λmax 214, 294 nm |
Peak Retention Time | 5.7 min |
Description
Class of Compounds | Drug |
Analyzing Compounds | Sildenafil or Viagra |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A