High Performance Liquid Chromatography (HPLC) Method for Analysis of Ammonium Ion, Lithium, Potassium, Sodium on Primesep 100 by SIELC Technologies
Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of Ammonium Ion, Lithium, Potassium, Sodium
Inorganic cations and ammonia are usually analyzed by ion-exchange chromatography with a conductivity detector. Mixed-mode chromatography achieves the same goal with an ELSD detector. In addition to retaining ions, mixed-mode column can retain hydrophobic molecules also, making it possible to analyze a wide variety of ionic and non-ionic compounds in one run. In this method sodium, potassium, and lithium ions were separated along with ammonium ions on a Primesep 100 column. Detection technique is ELSD.
Column | Primesep 100, 4.6 x 250 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 80/20% |
Buffer | TFA – 0.1% |
Flow Rate | 1.0 ml/min |
Detection | ELSD, the nebulizer and evaporator temperatures 50°C, with a gas flow rate of 1.6 Standard Liters per Minute (SLM) |
Class of Compounds | Ions, Hydrophilic, Ionizable, Vitamin, Supplements |
Analyzing Compounds | Sodium, Potassium, Lithium, Ammonium |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 250 mm
Particle Size: 5 µm
Pore Size: 100 A
Lithium
Potassium
Sodium