HPLC Method for Tetracycline on Primesep 100 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Tetracycline
Tetracycline is an antibiotic with the molecular formula C22H24N2O8. It is a bacteriostatic agent, therefore it prevents bacteria from growing and multiplying through inhibiting protein synthesis within bacteria. It is used to treat a variety of infections.
Tetracycline can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a sulfuric acid buffer. Detection is performed using UV.
Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | Gradient MeCN – 20-80% |
Buffer | H2SO4 – 0.2% |
Flow Rate | 1.0 ml/min |
Detection | UV 275 nm |
Class of Compounds | Medication |
Analyzing Compounds | Tetracycline |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended