HPLC Method for N-Nitrosodiethylamine on Newcrom R1 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of N-Nitrosodiethylamine
N-Nitrosodiethylamine (NDEA) or Diethylnitrosamine is potential carcenogenic agent was found in some blood pressure medications. Those medications were recalled by FDA with the concern of high concentration of the chemical.
SIELC developed simple and fast HPLC method to measure presence of NDEA in drug formulations. This UV based method allows to detect ppb level of the compound in solution. Simple mobile phase comprised of acetonitrile (MeCN) and water with detection at 2** nm using Newcrom R1 column can be used to detect both the API and the NDEA.
Column | Newcrom R1, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | MeCN – 20% |
Buffer | No |
Flow Rate | 1.0 ml/min |
Detection | UV, 232 nm |
Class of Compounds | Carcinogen, Zwitterionic, Hydrophobic |
Analyzing Compounds | N-Nitrosodiethylamine |
Application Column
Newcrom R1
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended