HPLC Method for 4-Ethoxyaniline on Primesep 100 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of 4-Ethoxyaniline.
4-Ethoxyaniline, also known as p-Phenetidine, is one of three Phenetidine isomers, with the chemical formula C8H11NO. It is commonly used as an intermediate product in drug and dye synthesis. You can find detailed UV spectra of 4-Ethoxyaniline and information about its various lambda maxima by visiting the following link.
4-Ethoxyaniline can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a Sulfuric Acid buffer. Detection is performed using UV.
| Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN – 40% |
| Buffer | H2SO4 – 0.2% |
| Flow Rate | 1.0 ml/min |
| Detection | UV 200nm |
| Class of Compounds | Aromatic amine |
| Analyzing Compounds | 4-Ethoxyaniline |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended