HPLC Method for Cyproheptadine on Primesep 100 by SIELC Technologies

Cyproheptadine is a first-generation antihistamine with the formula C21H21N. It is used primarily in treatment of allergies, insomnia, and serotonin syndrome. It is occasionally used as an appetite stimulant. It works sas the inverse agonist of the histamine H1 receptor. You can find detailed UV spectra of Cyproheptadine and information about its various lambda maxima by visiting the following link.
Cyproheptadine can be retained, and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water, acetonitrile (MeCN), and H2SO4 as a buffer. Detection is carried out using UV
| Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN/H2O – 70/30% |
| Buffer | H2SO4 – 0.1% |
| Flow Rate | 1 ml/min |
| Detection | UV 200 nm |
| LOD* | 5 ppb |
| Class of Compounds | Medication |
| Analyzing Compounds | Cyproheptadine |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended