High Performance Liquid Chromatography (HPLC) Method for Analysis of Hymexazol on Primesep B by SIELC Technologies
Separation type: Liquid Chromatography Mixed-mode SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Hymexazol
3-Hydroxy-5-methylisoxazole can be retained and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV at 222 nm
Column | Primesep B, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 10/90% |
Buffer | H2SO4 – 0.05% |
Flow Rate | 1.0 ml/min |
Detection | UV 220 nm |
Samples | 1.0 mg/ml in MeCN/H2O – 50/50% |
Injection volume | 1 µl |
LOD* | 7 ppb |
Class of Compounds | Isoxazoles |
Analyzing Compounds | Hymexazol |
Application Column
Primesep B
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Application Analytes:
HymexazolApplication Detection:
UV Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.