High Performance Liquid Chromatography (HPLC) Method for Analysis of Hymexazol on Primesep B by SIELC Technologies
Separation type: Liquid Chromatography Mixed-mode SIELC Technologies

High Performance Liquid Chromatography (HPLC) Method for Analysis of Hymexazol
3-Hydroxy-5-methylisoxazole can be retained and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV at 222 nm
Column | Primesep B, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
Mobile Phase | MeCN/H2O – 10/90% |
Buffer | H2SO4 – 0.05% |
Flow Rate | 1.0 ml/min |
Detection | UV 220 nm |
Samples | 1.0 mg/ml in MeCN/H2O – 50/50% |
Injection volume | 1 µl |
LOD* | 7 ppb |
Class of Compounds | Isoxazoles |
Analyzing Compounds | Hymexazol |
Application Column
Primesep B
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended