HPLC Method for Analysis of Guaiacol on Primesep 100 by SIELC Technologies
Separation type: Liquid Chromatography Mixed-mode
Guaiacol is an organic compound with the chemical formula C7H8O2. It is a naturally occurring substance derived from guaiacum or wood creosote, and it is known for its characteristic smoky odor.
Structure and Properties:
Guaiacol is a phenolic compound, specifically a methoxyphenol.
It appears as a colorless to pale yellow aromatic oil.
It has a molecular structure consisting of a benzene ring with methoxy (-OCH3) and hydroxyl (-OH) groups attached.
Sources:
Naturally found in guaiacum wood, creosote, and certain plants.
Can be produced synthetically through the methylation of catechol or other methods.
Uses:
Flavor and Fragrance: Guaiacol contributes to the smoky flavor in various foods and is used in flavorings and perfumes.
Medical Applications: It has been used in expectorants and antiseptics. Guaiacol derivatives are also used in the pharmaceutical industry.
Chemical Intermediate: It serves as a precursor for the synthesis of other chemicals, such as vanillin.
Health and Safety:
As a natural compound, guaiacol is biodegradable.
Its release into the environment through industrial processes should be managed to prevent ecological damage.
Guaiacol’s unique properties make it valuable in various industries, from food and fragrance to pharmaceuticals and chemical manufacturing.
Guaiacol can be retained and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV at 210 nm
High Performance Liquid Chromatography (HPLC) Method for Analysis of Guaiacol
Condition
Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 10/90% |
Buffer | H2SO4 – 0.2% |
Flow Rate | 1.0 ml/min |
Detection | UV 210, 275 nm |
Peak Retention Time | 4.21 min |
Description
Class of Compounds | Phenols |
Analyzing Compounds | Guaiacol |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A