HPLC Method for 6-Methyl-5-hepten-2-one on Newcrom R1 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of 6-Methyl-5-hepten-2-one.
Sulcatone is a ketone used as a mosquito attractant. It is a volatile oil component found in
citrus. It can be retained on a Newcrom R1reverse-phase column with a mobile phase of acetonitrile (ACN) and water with sulfuric acid (H2SO4) buffer. The analytical method is MS-compatible and can also use UV detection at 200nm.
Column | Newcrom R1 |
Mobile Phase | MeCN/H2O – 50/50% |
Buffer | H2SO4 – 0.1 % |
Flow Rate | 1.0 ml/min |
Detection | UV, 200 nm |
Class of Compounds |
Hydrophobic, Mosquito attractant |
Analyzing Compounds | 6-Methyl-5-hepten-2-one |
Application Column
Newcrom R1
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended