HPLC Method for Blasticidin-S on BIST B by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Blasticidin-S
Blasticidin S is an antibiotic with the chemical formula C17H26N8O5. It is primarily used in biology research for selecting cells of interest among cell cultures. It works through disrupting protein translation within human cells, fungi, and bacteria due to it’s resemblance to cytidine. You can find detailed UV spectra of Blasticidin-S and information about its various lambda maxima by visiting the following link.
Blasticidin-S can be retained and analyzed using the BIST B stationary phase column. The analysis utilizes a gradient method with a simple mobile phase consisting of acetonitrile (MeCN). Detection is performed using UV.
| Column | BIST B, 4.6 x 50 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | Gradient MeCN/H2O – 80/20 – 50/50 %, 10 min |
| Buffer | Sulfuric Acid – 0.2% |
| Flow Rate | 1.0 ml/min |
| Detection | UV 275nm |
| Limit of Detection* | 13 ppb |
| Class of Compounds | Antibiotic |
| Analyzing Compounds | Blasticidin-S |
Application Column
BIST B
Column Diameter: 4.6 mm
Column Length: 50 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended