| CAS Number | 528-48-3 |
|---|---|
| Molecular Formula | C15H10O6 |
| Molecular Weight | 286.239 |
| InChI Key | XHEFDIBZLJXQHF-UHFFFAOYSA-N |
| LogP | 2 |
| Synonyms |
|
Applications:
Uv-Vis Spectrum of Fisetin
February 23, 2026
Access the UV-Vis Spectrum SIELC Library

If you are looking for optimized HPLC method to analyze Fisetin check our HPLC Applications library
For optimal results in HPLC analysis, it is recommended to measure absorbance at a wavelength that matches the absorption maximum of the compound(s) being analyzed. The UV spectrum shown can assist in selecting an appropriate wavelength for your analysis. Please note that certain mobile phases and buffers may block wavelengths below 230 nm, rendering absorbance measurement at these wavelengths ineffective. If detection below 230 nm is required, it is recommended to use acetonitrile and water as low UV-transparent mobile phases, with phosphoric acid and its salts, sulfuric acid, and TFA as buffers.
For some compounds, the UV-Vis Spectrum is affected by the pH of the mobile phase. The spectra presented here are measured with an acidic mobile phase that has a pH of 3 or lower.
HPLC Method for Analysis of Fisetin
October 10, 2018
HPLC Method for Fisetin on Newcrom R1 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Fisetin
Fisetin, is a plant polyphenol from the flavonoid group. It can be found in many plants, where it serves as a coloring agent. It is also found in many fruits and vegetables, such as strawberries, apples, persimmons, onions and cucumbers. Fisetin is a natural flavonoid with various benefits, such as antioxidant, anticancer, neuroprotection effects.
Fisetin can be retained and analyzed using the Newcrom R1 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with a phosphoric acid base. Detection is performed using UV.
| Column | Newcrom R1, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN/H2O – 25/75% |
| Buffer | H3PO4 – 0.1% |
| Flow Rate | 1.0 ml/min |
| Detection | UV, 250 nm |
| Class of Compounds |
Hydrophobic, Ionizable, Coloring agent, Polyphenol |
| Analyzing Compounds | Fisetin |
Application Column
Newcrom R1
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended