CAS Number | 3922-86-9 |
---|---|
Molecular Formula | C9H20NO2+ |
Molecular Weight | 174.26 |
InChI Key | YRIBGSCJIMXMPJ-UHFFFAOYSA-N |
LogP | 1 |
Synonyms |
|
Applications:
HPLC ELSD Method for Analysis Butyrylcholine on Obelisc R Column
January 25, 2024
HPLC Method for Analysis of Butyrylcholine on Obelisc R by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Butyrylcholine
Butyrylcholine can be retained and analyzed using an Obelisc R mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase consisting of water, acetonitrile (MeCN), and ammonium formate as a buffer. Detection is achieved using ELSD
Column | Obelisc R, 2.1 x 100 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 70/30% |
Buffer | Ammonium formate pH 3.0 – 2 mM |
Flow Rate | 0.2 ml/min |
Detection | ELSD, the nebulizer and evaporator temperatures 50°C, with a gas flow rate of 1.6 Standard Liters per Minute (SLM) |
Samples | 1 mg/ml |
Injection volume | 1 µl |
LOD* | 60 ppb |
Class of Compounds | Quaternity amines |
Analyzing Compounds | Butyrylcholine |
Application Column
Obelisc R
Column Diameter: 2.1 mm
Column Length: 100 mm
Particle Size: 5 µm
Pore Size: 100 A
Application Analytes:
ButyrylcholineApplication Detection:
ELSD Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.