| CAS Number | 235114-32-6 |
|---|---|
| Molecular Formula | C56H71N9O23S |
| Molecular Weight | 1270.3 |
| InChI Key | PIEUQSKUWLMALL-YABMTYFHSA-N |
| LogP | 1.5 |
| Synonyms |
|
Applications:
Uv-Vis Spectrum of Micafungin
April 8, 2026
Access the UV-Vis Spectrum SIELC Library

If you are looking for optimized HPLC method to analyze Micafungin check our HPLC Applications library
For optimal results in HPLC analysis, it is recommended to measure absorbance at a wavelength that matches the absorption maximum of the compound(s) being analyzed. The UV spectrum shown can assist in selecting an appropriate wavelength for your analysis. Please note that certain mobile phases and buffers may block wavelengths below 230 nm, rendering absorbance measurement at these wavelengths ineffective. If detection below 230 nm is required, it is recommended to use acetonitrile and water as low UV-transparent mobile phases, with phosphoric acid and its salts, sulfuric acid, and TFA as buffers.
For some compounds, the UV-Vis Spectrum is affected by the pH of the mobile phase. The compound was dissolved with acetonitrile. This analysis was done with the mobile phase made of 20% acetonitrile and 80% water.
HPLC Method for Determination of Micafungin on Newcrom R1 Column
September 12, 2022
HPLC Method for Micafungin on Newcrom R1 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Micafungin
Micafungin is an anti-fungal medication with the chemical formula C56H71N9O23S. It is used to treat infections including but not limited to candidemia, abscesses, and esophageal candidiasis. It works through inhibiting the production of beta-1,3-glucan. It is only accessible through perscription.
Micafungin can be retained and analyzed on a Newcrom R1 reverse-phase column using a simple isocratic analytical method with a mobile phase consisting of Acetonitrile (MeCN), water, and Sulfuric acid as the ionic modifier. This analysis method can be UV detected at 270 nm with high resolution and peak symmetry.
| Column | Newcrom R1, 2.1 x 100 mm, 3 µm, 100 A, dual ended |
| Mobile Phase | MeCN – 45% |
| Buffer | H2SO4 – 0.2% |
| Flow Rate | 0.2 ml/min |
| Detection | UV 270 nm |
| LOD | 0.02 ppm |
| Class of Compounds | Drug |
| Analyzing Compounds | Micafungin |
Application Column
Newcrom R1
Column Diameter: 2.1 mm
Column Length: 100 mm
Particle Size: 3 µm
Pore Size: 100 A
Column options: dual ended