HPLC Separation of Growth Regulators (Paclobutrazol, Uniconazole, Daminozide) on Newcrom A Column


HPLC.cloud
View on hplc.cloud

 

 

 


HPLC.cloud
View on hplc.cloud

Condition 1

Column Newcrom A, 4.6×150 mm, 5 µm, 100A
Mobile Phase MeCN/H2O
Buffer Formic Acid
Flow Rate 1.0 ml/min
Detection CAD, UV 268 nm

 

Condition 2

Column Newcrom A, 4.6×150 mm, 5 µm, 100A
Mobile Phase MeCN/H2O
Buffer H2SO4
Flow Rate 1.0 ml/min
Detection UV 210 nm

Description

Class of Compounds Hydrophobic, Herbicide, Pesticide
Analyzing Compounds Paclobutrazol, Uniconazole, Daminozide

Application Column

Newcrom A

The Newcrom columns are a family of reverse-phase-based columns. Newcrom A, AH, B, and BH are all mixed-mode columns with either positive or negative ion-pairing groups attached to either short (25 Å) or long (100 Å) ligand chains. Newcrom R1 is a special reverse-phase column with low silanol activity.

Select options
Application Analytes:
Daminozide
Paclobutrazol
Uniconazole-P
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.