HPLC Method for Analysis of Sucrose Stearate on Chromni Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Sucrose Stearate.
Sucrose Stearate is a sugar ester with the chemical formula C30H58O13. It is used primarily as an emulsifier and moisturizer in lotions, creams, and cleansers.
Sucrose Stearate can be retained and analyzed using the Chromni stationary phase column. The analysis utilizes a gradient method with a simple mobile phase consisting of ethanol and acetonitrile (MeCN). Detection is performed using ELSD.
Condition
| Column | Chromni, 3.2 x 100 mm, 3 µm, 100 A, dual ended |
| Mobile Phase | MeCN/EtOH |
| Buffer | None |
| Flow Rate | 1.0 ml/min |
| Detection | ELSD, the nebulizer and evaporator temperatures 50°C with a gas flow rate of 1.6 Standard Liters per Minute (SLM) |
Description
| Class of Compounds | Sugar |
| Analyzing Compounds | Sucrose Stearate |
Application Column
Chromni
Column Diameter: 3.2 mm
Column Length: 100 mm
Particle Size: 3 µm
Pore Size: 100 A
Column options: dual ended