HPLC Method for Myelin Basic Protein Peptide Antagonist on BIST B+ by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Myelin Basic Protein Peptide Antagonist
Myelin Basic Protein Peptide Antagonist is a synthetic peptide antagonist with the molecular formula C70H114N18O21 • XCF3COOH. When expressing an MBP (85-99), it inhibits the production of IL-2 in T cell hybridoma cells. MBP peptide antagonist also delays MBP (85-99)-induced experimental autoimmune encephalomyelitis and reduces symptom severity.
Myelin Basic Protein Peptide Antagonist can be retained and analyzed using the BIST B+ stationary phase column. The analysis utilizes a gradient method with a simple mobile phase consisting of water and acetonitrile (MeCN). Detection is performed using UV.
| Column | BIST B+, 4.6 x 50 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | Gradient MeCN – 70/30% – 50/50%, 10 min |
| Buffer | Sulfuric Acid |
| Flow Rate | 1.0 ml/min |
| Detection | UV 200 |
| LOD* UV | 194 ppb |
| Class of Compounds | Peptide Antagonist |
| Analyzing Compounds | Myelin Basic Protein Peptide Antagonist |
Application Column
BIST B+
Column Diameter: 4.6 mm
Column Length: 50 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended