| CAS Number | 2032-65-7 |
|---|---|
| Molecular Formula | C11H15NO2S |
| Molecular Weight | 225.310 |
| InChI Key | YFBPRJGDJKVWAH-UHFFFAOYSA-N |
| LogP | 2.92 |
| Synonyms |
|
Applications:
HPLC Method for Analysis of Methiocarb on Newcrom B Column on Alltesta™
January 8, 2026
HPLC Method for Methiocarb on Newcrom B by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Methiocarb
Methiocarb is a carbamate pesticide with the chemical formula C11H15NO2S. It works as a acetylcholinesterase inhibitor. It has been in use as an insecticide, bird repellent, acaricide, and molluscicide since 1960s.
Methiocarb can be retained and analyzed using the Newcrom B stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with ammonium formate as a buffer. Detection is performed using UV.
| Column | Newcrom B, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN – 50% |
| Buffer | Ammonium Formate |
| Flow Rate | 1.0 mL/min |
| Detection | UV 275 nm |
| Class of Compounds | Pesticide |
| Analyzing Compounds | Methiocarb |
Application Column
Newcrom B
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
UV-Vis Spectrum of Methiocarb
December 23, 2025
If you are looking for optimized HPLC method to analyze Methiocarb check our HPLC Applications library
For optimal results in HPLC analysis, it is recommended to measure absorbance at a wavelength that matches the absorption maximum of the compound(s) being analyzed. The UV spectrum shown can assist in selecting an appropriate wavelength for your analysis. Please note that certain mobile phases and buffers may block wavelengths below 230 nm, rendering absorbance measurement at these wavelengths ineffective. If detection below 230 nm is required, it is recommended to use acetonitrile and water as low UV-transparent mobile phases, with phosphoric acid and its salts, sulfuric acid, and TFA as buffers.
For some compounds, the UV-Vis Spectrum is affected by the pH of the mobile phase. The spectra presented here are measured with an acidic mobile phase that has a pH of 3 or lower.
Separation of Methiocarb on Newcrom R1 HPLC column
February 16, 2018
Methiocarb can be analyzed by this reverse phase (RP) HPLC method with simple conditions. The mobile phase contains an acetonitrile (MeCN), water, and phosphoric acid. For Mass-Spec (MS) compatible applications the phosphoric acid needs to be replaced with formic acid. Smaller 3 µm particles columns available for fast UPLC applications. This liquid chromatography method is scalable and can be used for isolation impurities in preparative separation. It also suitable for pharmacokinetics.
Application Column
Newcrom R1
The Newcrom columns are a family of reverse-phase-based columns. Newcrom A, AH, B, and BH are all mixed-mode columns with either positive or negative ion-pairing groups attached to either short (25 Å) or long (100 Å) ligand chains. Newcrom R1 is a special reverse-phase column with low silanol activity.