HPLC MS Method for Analysis of Amoxicillin on Primesep 100 Column

HPLC MS Method for Analysis of Procainamide on Primesep 100 Column by SIELC Technologies

HPLC MS Method for Analysis of Amoxicillin on Primesep 100 Column
HPLC Method for Analysis of Amoxicillin on Primesep 100 Column by SIELC Technologies

Amoxicillin is a prescription antibiotic with the molecular formula C16H19N3O5S. It is typically used to treat bacterial infections and is taken orally. It belongs to the penicillin class. It is used to treat a variety of bacterial infections, including respiratory tract infections, ear infections, skin infections, urinary tract infections, and certain types of bacterial endocarditis. Amoxicillin works by inhibiting the synthesis of bacterial cell walls. It does this by interfering with the formation of peptidoglycan, an essential component of the bacterial cell wall. It is effective against a broad spectrum of bacteria, including both Gram-positive and Gram-negative bacteria. However, its effectiveness may be limited against bacteria that produce beta-lactamase enzymes.

Amoxicillin can be retained and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and ammonium formate as a buffer. This method allows for detection using UV at 272 nm

ColumnPrimesep 100, 2.1 x 100 mm, 5 µm, 100 A, dual ended
Mobile PhaseMeCN 20%
BufferAmFm pH 3.0  – 15 mM
Flow Rate0.2 ml/min
DetectionUV 272nm, SIM + 349, 366
Samples1 mg/ml
Injection volume5 µl
LOD*0.3 ppm
* LOD was determined for this combination of instrument, method, and analyte, and it can vary from one laboratory to another even when the same general type of analysis is being performed.
Class of Compounds
Drug
Analyzing CompoundsAmoxicillin

Application Column

Primesep 100

Column Diameter: 2.1 mm
Column Length: 100 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended

Add to cart
Application Analytes:
Amoxicillin

Application Detection:
UV Detection
LC MS Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.