HPLC Method for Analysis of Ataciguat on Chromni Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Ataciguat.
Ataciguat is an drug with the chemical formula C21H19N3O6S3. It is primarily used as a guanylate cyclase activator. Through activating sGC and restoring the NO-cGMP pathways, it helps to reduce calcium deposition, slow valve calcification, and generally treats calcific aortic valve stenosis. t helps reduce calcium deposition You can find detailed UV spectra of Ataciguat and information about its various lambda maxima by visiting the following link.
Ataciguat can be retained and analyzed using the Chromni stationary phase column. The analysis utilizes a gradient method with a simple mobile phase consisting of water and acetonitrile (MeCN) and ethanol. Detection is performed using UV.
Condition
| Column | Chromni, 4.6 x 150 mm, 3 µm, 100 A, dual ended |
| Mobile Phase | MeCN/H2O -30/70 – 45/55%, 10 min |
| Buffer | AmAc pH 5.0 – 10 mM |
| Flow Rate | 1.0 ml/min |
| Detection | UV 270 nm |
| Limit Of Detection* | 24.5 ppb |
Description
| Class of Compounds | Drug |
| Analyzing Compounds | Ataciguat |
Application Column
Chromni
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 3 µm
Pore Size: 100 A
Column options: dual ended




