| CAS Number | 83602-39-5 |
|---|---|
| Molecular Formula | C62H111N11O12 |
| Molecular Weight | 1202.6 |
| InChI Key | PMATZTZNYRCHOR-JLPRAAIDSA-N |
| LogP | 7.5 |
| Synonyms |
|
Applications:
HPLC Method for Analysis of Cyclosporine H on Legacy L16 Column
September 2, 2026
HPLC Method for Analysis of Cyclosporin H on Legacy L16 by SIELC Technologies

Cyclosporine H is a selective inhibitor of formyl peptide receptor-1 (FPR-1) with the molecular formula C61H111N11O12. It is used primarily used in research. Unlike Cyclosporin A, B, and C, it does not have immunosuppressive properties. You can find detailed UV spectra of Cyclosporine H and information about its various lambda maxima by visiting the following link.
Cyclosporin H can be retained and analyzed using the Legacy L16 stationary phase column. The analysis utilizes an isocratic method with a mobile phase consisting of acetonitrile, water, methanol, and phosphoric acid. Detection is performed using UV.
Condition
| Column | Legacy L16, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN/MeOH/H2O – 50/5/45% |
| Buffer | H3PO4 – 0.05% |
| Flow Rate | 1.0 ml/min |
| Detection | UV 210 |
| Column Temperature | 70 C |
Description
| Class of Compounds | Inhibitor |
| Analyzing Compounds | Cyclosporin H |
Application Column
Legacy L16
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
Uv-Vis Spectrum of Cyclosporin H
March 16, 2026
Access the UV-Vis Spectrum SIELC Library

If you are looking for optimized HPLC method to analyze Cyclosporin H check our HPLC Applications library
For optimal results in HPLC analysis, it is recommended to measure absorbance at a wavelength that matches the absorption maximum of the compound(s) being analyzed. The UV spectrum shown can assist in selecting an appropriate wavelength for your analysis. Please note that certain mobile phases and buffers may block wavelengths below 230 nm, rendering absorbance measurement at these wavelengths ineffective. If detection below 230 nm is required, it is recommended to use acetonitrile and water as low UV-transparent mobile phases, with phosphoric acid and its salts, sulfuric acid, and TFA as buffers.
For some compounds, the UV-Vis Spectrum is affected by the pH of the mobile phase. The compound was dissolved in acetonitrile.