Alltesta HPLC Method for Polyethylenimine PEI on PEI by SIELC Technologies
Separation type: Liquid Chromatography Reversed-phase
Polyethyleneimine (PEI) is a polymer used in a wide range of applications. It is difficult to analyze in HPLC due to several factors. It’s a mixture of compounds with different lengths and therefore different number of charges. It also lacks a UV chromophore. New PEI column was designed specifically for the analysis of polyethyleneimine by ion-exclusion and size-exclusion mechanisms with copper complex to allow for UV detection. The method uses a mobile phase of acetonitrile (ACN) and water with ammonium formate buffer (AmFm) and UV detection at 275nm
High Performance Liquid Chromatography (HPLC) Method for Analyses of Polyethylenimine PEI on PEI Column by SIELC Technologies
Condition
Column | PEI, 4.6 x 250 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 40/60%, |
Buffer | Ammonium formate pH 3.0 – 20 mM |
Flow Rate | 0.5 ml/min |
Detection | UV 275 nm |
Description
Class of Compounds | Polymer, Basic, Hydrophilic, Ionizable |
Analyzing Compounds | Polyethylenimine PEI |