HPLC Method for Analysis of Insulin on BIST™B+ Column
July 8, 2022
Separation type: Bridge Ion Separation Technology, or BIST™
High Performance Liquid Chromatography (HPLC) Method for Analysis of Insulin.
Insulin is an anabolic peptide hormone naturally produced in the pancreas that regulates the body’s glucose levels. It is also synthesized as a treatment drug for those with Diabetes who can’t regulate their own blood sugar levels naturally. Two well-known forms of synthetic insulin are NovoLog and Lantus. NovoLog is a fast-acting version of insulin and can be used by both Type-1 and Type-2 Diabetes patients, whereas Lantus is a long-acting version of insulin that is primarily used by Type-2 Diabetes patients (and by Type-1 patients under the age of 6).
Using SIELC’s newly introduced BIST™ method, these two forms of insulin can be separated and retained on a positively-charged anion-exchange BIST™B column. There are two keys to this retention method: 1) a multi-charged, negative buffer, such as Sulfuric acid (H2SO4), which acts as a bridge, linking the positively-charged amine analytes to the positively-charged column surface and 2) a mobile phase consisting mostly of organic solvent to minimize the formation of a solvation layer around the charged analytes. This method actually utilizes a gradient that lowers the MeCN concentration over time; this reduces the retention effects on the insulin analytes, allowing for quicker elution times while still maintaining clear selectivity and separation. Using this new and unique analysis method, two different manu can be retained and UV detected at 210 nm.
| Column | BIST B+, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN Gradient |
| Buffer | H2SO4 – 0.2% |
| Flow Rate | 1.0 ml/min |
| Detection | UV 210 nm |
| Class of Compounds |
Polymer, Protein |
| Analyzing Compounds | Insulin, Insulin, neutral |
Application Column
BIST B+
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
Insulin, neutral
HPLC Separation of Human and Synthetic Insulins on Promix Column
May 6, 2011

Insulin is a protein that regulates carbohydrates and fat metabolism. Insulin is a peptide hormone consisting of 51 amino acids and has molecular weight of 5808 Da. Bovine insulin differs from human in only three amino acid residues, and porcine insulin in one. Difference in structures is very minimal and the insulin cannot be resolve on reversed-phase column. Mixed-mode approach allows to use small difference in hydrophobic and ionic properties and separate insulins with close structures. Normal human insulin, synthetic insulin Humalog, and synthetic insulin Lantus are separated on a Promix MP mixed-mode HPLC column with LC/MS-compatible conditions. Method does not require use of ion-pairing reagent. Method is also compatible with preparative chromatography and can be used for isolation of insulin on a large scale.
Application Column
Promix MP
The Promix family of mixed-mode columns presents an innovative chromatography technology for the efficient resolution of peptides and proteins. This technology hinges on a unique blend of hydrophobic and ionic interactions, facilitated by a novel separation medium: a ligand bonded to a silica support, chemically combined with hydrophobic and ionic functional groups. This phase provides unparalleled selectivity and peak capacity. By independently adjusting the quantities of buffer and organic modifier, a virtually infinite number of separation conditions can be achieved, rendering it suitable for a wide array of biomolecules.
Select optionsInsulin
Lantus