Folpet is a nonsystemic fungicide which is used to prevent apple scab, cherry leaf spot, rose black spot, and rose mildew. A method for folpet was provided by the EURL-SRM, a proficiency test for sigle residue methods developed by the European Union Reference Laboratory. Folpet was retained and separated from impurities with Obelisc R and Primesep B2 columns. Primesep B2 contains embedded basic ion-pairing groups and Obelisc R contains embedded ionic and hydrophobic groups which can assist in fine tuning separations. Method can be used for dozens of other pesticides and is LC/MS compatible.
Column | Primesep B2, 2.1 x 150 mm, 5 µm, 100 A |
Mobile Phase | Gradient MeCN – 10-70%, 15 min, 7 min hold |
Buffer | Gradient AmAc pH 4,5- 20-50 mM, 15 min, 7 min hold |
Flow Rate | 0.4 ml/min |
Detection | UV, 250 nm |
Class of Compounds |
Insecticide, Herbicide, Fungicide, Hydrophobic, Ionizable |
Analyzing Compounds | Folpet |
Application Column
Primesep B2
Column Diameter: 2.1 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Application Analytes:
FolpetApplication Detection:
UV Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.