| CAS Number | 163521-12-8 |
|---|---|
| Molecular Formula | C26H27N5O2 |
| Molecular Weight | 441.535 |
| InChI Key | SGEGOXDYSFKCPT-UHFFFAOYSA-N |
| LogP | 4 |
| Synonyms |
|
Applications:
HPLC Method for Analysis of Vilazodone in Pharmaceutical Formulation on Primesep 100 Column
August 11, 2026
HPLC Method for Analysis of Vilazodone on Primesep 100 Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Vilazodone.
Vilazodone is a serotonin modulator with the chemical formula C26H27N5O2. It is primarily used to treat major depressive disorder. You can find detailed UV spectra of Vilazodone and information about its various lambda maxima by visiting the following link.
Vilazodone can be retained and analyzed using the Primesep 100 stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN). Detection is performed using UV.
Condition
| Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN/H2O – 50/50 |
| Buffer | H2SO4 – 0.2% |
| Flow Rate | 1.0 ml/min |
| Detection | UV 240 nm |
| Limit Of Detection* | 80 ppb |
Description
| Class of Compounds | Drug |
| Analyzing Compounds | Vilazodone |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended
Uv-Vis Spectrum of Vilazodone
August 5, 2026
Access the UV-Vis Spectrum SIELC Library

If you are looking for optimized HPLC method to analyze Vilazodone check our HPLC Applications library
UV/Vis.: λmax: 202, 240, 372, 313 nm
For optimal results in HPLC analysis, it is recommended to measure absorbance at a wavelength that matches the absorption maximum of the compound(s) being analyzed. The UV spectrum shown can assist in selecting an appropriate wavelength for your analysis. Please note that certain mobile phases and buffers may block wavelengths below 230 nm, rendering absorbance measurement at these wavelengths ineffective. If detection below 230 nm is required, it is recommended to use acetonitrile and water as low UV-transparent mobile phases, with phosphoric acid and its salts, sulfuric acid, and TFA as buffers.
For some compounds, the UV-Vis Spectrum is affected by the pH of the mobile phase. The compound was dissolved in ethanol. This analysis was done with the mobile phase of hexane, IPA, and TFA.
This compound was diluted in a mixture of water and acetonitrile. The method was done with a mixture of water, acetonitrile, and a sulfuric acid buffer.