HPLC Method for Analysis of Metalaxyl on Cyralis CM Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Metalaxyl.
Metalaxyl, also known as methyl N-(methoxyacetyl)-N-(2,6-xylyl)-DL-alaninate, is a systemic fungicide with the chemical formula C15H21NO4. It is used to control Pythium and Phytophthora in vegetables. It has suffered a resistance problem going back to the 1980s.
You can find detailed UV spectra of Metalaxyl and information about its various lambda maxima by visiting the following link.
This application describes a normal-phase chiral HPLC method for the separation of Metalaxyl enantiomers using SIELC’s Cyralis CM chiral column. Cyralis CM is a cellulose-based chiral stationary phase (CSP) designed for enantiomeric separations, optical purity testing, chiral impurity analysis, and stereoisomer characterization in pharmaceutical, chemical, agrochemical, and fine chemical applications.
Metalaxyl can be retained and analyzed using the Cyralis CM stationary phase column. The analysis utilizes an isocratic method with a mobile phase consisting of Ethanol. Detection is performed using UV.
Condition
| Column | Cyralis CM, 4.6 x 250 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | Ethanol |
| Flow Rate | 1.0 mL/min |
| Detection | UV 230 nm |
| Resolution | 8.40 |
Description
| Class of Compounds | Fungicide |
| Analyzing Compounds | Metalaxyl |
Application Column
Cyralis CM
Column Diameter: 4.6 mm
Column Length: 250 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended