HPLC Method for Analysis of Procymidone on Newcrom B Column on Alltesta™

HPLC Method for Procymidone on Newcrom B by SIELC Technologies


High Performance Liquid Chromatography (HPLC) Method for Analysis of Procymidone

Procymidone is a pesticide and fungicide with the chemical formula  C13H11Cl2NO2. It is typically used to kill unwanted ferns, nettles, and fungi in lupins, grapes, stone fruit, and strawberries.

Procymidone can be retained and analyzed using the Newcrom B stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) with phosphoric acid as a buffer. Detection is performed using UV.

ColumnNewcrom B, 4.6 x 150 mm, 5 µm, 100 A, dual ended
Mobile PhaseMeCN – 70%
BufferSulfuric Acid
Flow Rate1.0 mL/min
DetectionUV 275 nm
Class of CompoundsDicarboximide fungicides
Analyzing CompoundsProcymidone

Application Column

Newcrom B

Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended

Add to cart
Application Analytes:
Procymidone

Application Detection:
UV Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.