Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of Sulfosuccinic Acid.
Sulfosuccinic acid is a surfactant used in cosmetics and pharmaceuticals. Sulfosuccinic acid can be retained on the Newcrom BH mixed-mode column using an isocratic analytical method with a simple mobile phase of water, acetonitrile (MeCN, ACN), and sulphuric acid (H2SO4) buffer. The analysis method can be UV detected at 200 nm.
Column | Newcrom BH, 3.2×100 mm, 3 µm, 100A |
Mobile Phase | MeCN/H2O -10/90% |
Buffer | H2SO4 – 0.3% |
Flow Rate | 0.5 ml/min |
Detection | UV, 200 nm |
Class of Compounds |
Acid |
Analyzing Compounds | Sulfosuccinic Acid |
Application Column
Newcrom BH
The Newcrom columns are a family of reverse-phase-based columns. Newcrom A, AH, B, and BH are all mixed-mode columns with either positive or negative ion-pairing groups attached to either short (25 Å) or long (100 Å) ligand chains. Newcrom R1 is a special reverse-phase column with low silanol activity.