HPLC Determination of 1-pyridin-3-yl-ethylamine on Primesep 100 Column

HPLC Method for 1-pyridin-3-yl-ethylamine on Primesep 100 by SIELC Technologies


 

High Performance Liquid Chromatography (HPLC) Method for Analysis of 1-pyridin-3-yl-ethylamine.

1-pyridin-3-yl-ethylamine has the chemical formula C7H10N2.

1-pyridin-3-yl-ethylamine can be retained  on the Primesep 100 mixed-mode column using an isocratic analytical method with a simple mobile phase of water, acetonitrile (MeCN, ACN), and sulphuric acid (H2SO4) buffer. The analysis method can be UV detected at 260 nm.

Condition

Column Primesep 100, 4.6 x 150 mm, 5 µm, 100 A, dual ended
Mobile Phase MeCN/H2O – 20/80%
Buffer H2SO4 – 0.3%
Flow Rate 1.0 ml/min
Detection UV, 260 nm

 

Description

Class of Compounds
Amine
Analyzing Compounds 1-pyridin-3-yl-ethylamine

 

Application Column

Primesep 100

Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended

Add to cart
Application Analytes:
1-pyridin-3-yl-ethylamine

Application Detection:
UV Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.