HPLC Method for Analysis of Etodolac on Cyralis CM Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Etodolac.
Etodolac is a pyrenecarboxylic acid non-steroidal anti-inflammatory drug (NSAID) with the chemical formula C17H21NO3. It is used primarily to manage mild to moderate pain, fever, and inflammation through reducing the levels of prostaglandins.
This application describes a normal-phase chiral HPLC method for the separation of Etodolac enantiomers using SIELC’s Cyralis CM chiral column. Cyralis CM is a cellulose-based chiral stationary phase (CSP) designed for enantiomeric separations, optical purity testing, chiral impurity analysis, and stereoisomer characterization in pharmaceutical, chemical, agrochemical, and fine chemical applications. For best results, the sample should be prepared fresh for analysis.
Etodolac can be retained and analyzed using the Cyralis CM stationary phase column. The analysis utilizes an isocratic method with a mobile phase consisting of Hexane, IPA, and TFA. Detection is performed using UV.
Condition
| Column | Cyralis CM, 4.6 x 100 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | Hexane/IPA/TFA |
| Flow Rate | 1.0 mL/min |
| Detection | UV 230 nm |
| Resolution | 4.78 |
Description
| Class of Compounds | Drug |
| Analyzing Compounds | Etodolac |
Application Column
Cyralis CM
Column Diameter: 4.6 mm
Column Length: 100 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended