HPLC Chiral Method for Analysis of Dichlorprop Enantiomers on Cyralis AM Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Dichlorprop.
Dichlorprop is a chlorophenoxy herbicide with the chemical formula C9H8Cl2O3. It is used to kill annual and perennial broadleaf weeds. It is widely used within the United States. It works through increasing cell wall plasticity, biosynthesis of proteins, and the production of ethylene which leads to excessive cell division and therefore damage to vascular tissue. It is acutely toxic, but is a severe eye irritant. You can find detailed UV spectra of Dichlorprop and information about its various lambda maxima by visiting the following link.
This application describes a normal-phase chiral HPLC method for the separation of Dichlorprop enantiomers using SIELC’s Cyralis AM chiral column. Cyralis AM is an amylose-based chiral stationary phase (CSP) designed for enantiomeric separations, optical purity testing, chiral impurity analysis, and stereoisomer characterization in pharmaceutical, chemical, agrochemical, and fine chemical applications.
Dichlorprop can be retained and analyzed using the Cyralis AM stationary phase column. The analysis utilizes an isocratic method with a mobile phase consisting of Hexane, IPA, and TFA. Detection is performed using UV.
Condition
| Column | Cyralis AM, 4.6 x 250 mm, 5 µm, 100 A, normal phase |
| Mobile Phase | Hexane/IPA/TFA |
| Flow Rate | 1.0 mL/min |
| Detection | UV 230 nm |
| Resolution | 3.67 |
Description
| Class of Compounds | Fungicide |
| Analyzing Compounds | Dichlorprop |
Application Column
Cyralis AM
Column Diameter: 4.6 mm
Column Length: 250 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: normal phase