HPLC Method for Analysis of Malathion on Newcrom BH Column by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Malathion.
Malathion is the most widely used organophosphate pesticide with the chemical formula C10H19O6PS2. In the US, it is used heavily in residential landscaping and in public recreation areas, especially for fruit fly and mosquito eradication. While not considered toxic to humans, malathion metabolizes to a far more toxic malaoxon. The EURL (European Union Reference Laboratory), included malathion as a target pesticide for the EUPT-CF9 a proficiency test for cereals and feedingstuff that require muli-residue methods. can find detailed UV spectra of Malathion and information about its various lambda maxima by visiting the following link.
Malathion can be retained and analyzed using the Newcrom BH stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water and acetonitrile (MeCN) and ethanol. Detection is performed using LC MS.
Condition
| Column | Newcrom BH, 3.2 x 100 mm, 5 µm, 100 A, dual ended |
| Mobile Phase | MeCN/H2O – 40/60% |
| Buffer | AmFm pH 3.0 – 20 mM |
| Flow Rate | 0.5 ml/min |
| Detection | LC MS ESI SIM 331 [M+H]+ |
Description
| Class of Compounds | Pesticide |
| Analyzing Compounds | Malathion |
Application Column
Newcrom BH
Column Diameter: 3.2 mm
Column Length: 100 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended