High Performance Liquid Chromatography (HPLC) Method for Analysis of Labetalol on Primesep B by SIELC Technologies
Separation type: Liquid Chromatography Mixed-mode SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Labetalol
Labetalol is primarily used for the management of hypertension. It is classified as a beta-blocker, which works by blocking the effects of endogenous chemicals such as epinephrine on the heart and blood vessels. Through this mechanism, a reduction in heart rate, blood pressure, and myocardial strain is achieved.
Labetalol can be retained, separated and analyzed using a Primesep B mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV 230 nm.
You can find detailed UV spectra of Labetalol and information about its various lambda maxima by visiting the following link.
Column | Primesep B, 4.6 x 250 mm, 5 µm, 100 A |
Mobile Phase | MeCN – 15% |
Buffer | H2SO4 -0.2% |
Flow Rate | 0.7 ml/min |
Detection | UV 230 nm |
Class of Compounds | beta-blocker |
Analyzing Compounds | Labetalol |
Application Column
Primesep B
Column Diameter: 4.6 mm
Column Length: 250 mm
Particle Size: 5 µm
Pore Size: 100 A