Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of Gamma-aminobutyric acid (GABA), Vigabatrin, Phenibut, Pregabalin, Baclofen
All gabapentinoids can be retained and separated on the Primesep 100 mixed-mode column using an isocratic analytical method with a simple mobile phase of water, acetonitrile (MeCN, ACN), and Ammonium Formate (AmFm) buffer. Detection can be achieved with mass spectrometry (MS), evaporative light scattering detection (ELSD) and Charged aerosol detection (CAD).
Column | Primesep 100, 4.6×150 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 30/70% |
Buffer | AmFm pH 3.0 – 50 mM |
Flow Rate | 1.0 ml/min |
Detection | CAD (Corona) (MS-compatible mobile phase) |
Class of Compounds | Acid, Drug |
Analyzing Compounds | Gamma-aminobutyric acid (GABA), Vigabatrin, Phenibut, Pregabalin, Baclofen |
Application Column
Primesep 100
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select optionsPhenibut
Pregabalin
Vigabatrin
gamma-Aminobutyric Acid (GABA)