HPLC Separation of Catecholamines on Primesep 100 Column with Phosphate Buffers

chr_284.gif

The catecholamine neurotransmitters are amino-acid derivatives of tyrosine. DOPA, tyrosine, phenylalanine, norepinephrine, epinephrine, and dopamine and baseline are resolved on a Primesep 100 column with UV-transparent phosphate buffer. This method can be used for analysis of catecholamines and related impurities in various matrices. Peak order and retention time can be changed by changing the amount of ACN, buffer concentration and buffer pH. Various buffers can be used to accommodate desired detection technique. Primesep 100 is a reversed-phase cation-exchange mixed-mode column that can be used for analysis of polar neutral, polar ionizable, polar zwitter-ionic, hydrophobic neutral, and hydrophobic ionic compounds in the same run. Column can be operated in reverse-phase, cation-exchange, anion-exclusion, HILIC and mixed-modes depending on the mobile phase selection and nature of analytes. Column is compatible with LC/MS and does not require use of ion-pairing reagents.

Condition

Column Primesep 100, 4.6×150 mm, 5 µm, 100A
Mobile Phase MeCN/H2O
Buffer Na2HPO4
Flow Rate 1.0 ml/min
Detection UV, 210 nm

 

Description

Class of Compounds
Drug, Acid, Hydrophilic, Ionizable, Hormone
Analyzing Compounds Tyrosine, DOPA, Phenylalanine, Norepinephrine, Epinephrine, Dopamine

 

Application Column

Primesep 100

The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.

Select options
Application Analytes:
DOPA (3,4-dihydroxy-L-phenylalanine)
Dopamine
Epinephrine
Norepinephrine
Phenylalanine
Tyrosine

Application Detection:
UV Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.