Glutamine, valine, isoleucine, and leucine are baseline resolved in a dietary supplement with a mobile phase of acetonitrile, water, and sulfuric acid with ultraviolet (UV) detection at 210 nm on a Primesep 100 column. Increasing the sulfuric acid concentration allows elution of arginine. Primesep 100 separates underivatized amino acids and amines by combining reversed-phase and ion-exchange mechanisms. The HPLC separation uses a mobile phase of water, acetonitrile (MeCN, ACN), and sulfuric acid with ultraviolet (UV) detection at 210 nm.
Column | Primesep 100, 4.6×250 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 30/70% |
Buffer | H2SO4 |
Flow Rate | 1.0 ml/min |
Detection | UV, 210 nm |
Class of Compounds |
Drug, Acid, Hydrophilic, Ionizable, Vitamin, Supplements, Amino acid |
Analyzing Compounds | Arginine, Glutamine, L-Valine (Val/V), L-Isoleucine (Ile/I), L-Leucine (Leu/L) |
Application Column
Primesep 100
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.