High Performance Liquid Chromatography (HPLC) Method for Analysis of Isonicotinic Acid on Primesep 100 by SIELC Technologies
Separation type: Liquid Chromatography Mixed-mode SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Isonicotinic Acid
Isonicotinic Acid can be retained, separated and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs an isocratic method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV 200 nm.
You can find detailed UV spectra of Isonicotinic Acid and information about its various lambda maxima by visiting the following link.
Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN – 10% |
Buffer | H2SO4 -0.05% |
Flow Rate | 1.0 ml/min |
Detection | UV 200 nm |
Samples | 1.0 mg/ml in MeCN/H2O – 50/50% |
Injection volume | 1 µl |
LOD* | 2 ppb (200 nm) |
Class of Compounds | Acid |
Analyzing Compounds | Isonicotinic Acid |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A