Ritodrine (also Yutopar), is a drug used in relaxation of smooth muscle and to stop premature labor. It has been pulled from the US market due to its side effects. Ritodrine can be retained on a Primesep 100 column with great peak shape using isocratic method of 30/70 Acetonitrile (ACN) and water with 0.3% sulfuric acid as buffer. UV Detection 276nm.
Condition
Column Primesep 100, 4.6×150 mm, 5 µm, 100A Mobile Phase MeCN/H2O – 30/70% Buffer H2SO4 – 0.3 % Flow Rate 1.0 ml/min Detection UV, 276 nm
Description
Class of Compounds
Drug, Basic, Hydrophobic, Ionizable, Amine Analyzing Compounds Ritodrine
Application Column
Primesep 100
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select options