HPLC Determination of Perindopril on Primesep 200 column

HPLC Method for Perindopril on Primesep 200 by SIELC Technologies

HPLC Determination of Perindopril

High Performance Liquid Chromatography (HPLC) Method for Analysis of Perindopril.

Perindopril is an ACE inhibitor used in treating hypertension (high blood pressure). It comes in tablet form. Perindopril can be retained on a Primesep 200 column in under 8 minutes with a simple isocratic mobile phase and sulphuric acid buffer. Primesep 200 is a reverse-phase (RP) mixed-mode column with embedded weak acidic ion-pairing groups

 

Condition

Column Primesep 200
Mobile Phase MeCN/H2O – 20/80%
Buffer H2SO4 – 0.4 %
Flow Rate 1.0 ml/min
Detection UV, 205 nm

 

Description

Class of Compounds
Drug, Acid, Hydrophobic, Ionizable
Analyzing Compounds Perindopril

 

Application Column

Primesep 200

Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A
Column options: dual ended

Add to cart
Application Analytes:
Perindopril

Application Detection:
UV Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.