Alltesta HPLC Method for Furosemide on Primesep B4 by SIELC Technologies
High Performance Liquid Chromatography (HPLC) Method for Analysis of Furosemide
Furosemide is a loop diuretic commonly used to treat conditions like edema (fluid retention) and hypertension (high blood pressure).
Furosemide can be retained, and analyzed using a Primesep B mixed-mode stationary phase column. The analysis utilizes an isocratic method with a simple mobile phase consisting of water, acetonitrile (MeCN), and sulfuric acid as a buffer. Detection is achieved using UV at 275 nm
Column | Primesep B4, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN/H2O – 50/50% |
Buffer | H3PO4 -0.1% |
Flow Rate | 1.0 ml/min |
Detection | UV 275 nm |
Class of Compounds | Diuretic |
Analyzing Compounds | Furosemide |
You can view examples of chromatograms obtained using the Allesta instrument and Sielc columns by clicking here.
Application Column
Primesep B4
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A