CAS Number | 128196-01-0 |
---|---|
Molecular Formula | C20H21FN2O |
Molecular Weight | 324.4 |
InChI Key | WSEQXVZVJXJVFP-FQEVSTJZSA-N |
LogP | 3.2 |
Synonyms |
|
Applications:
HPLC Determination of Escitalopram (Cipralex) on Primesep SB Column
March 18, 2021
Separation type: Liquid Chromatography Mixed-mode
High Performance Liquid Chromatography (HPLC) Method for Analysis of Escitalopram
Escitalopram is a type of antidepressant known as a selective serotonin reuptake inhibitor. It is used to restore serotonergic function in the treatment of depression and anxiety. Escitalopram helps many people recover from depression. It can be retained on Primesep SB, a reverse-phase (RP) mixed-mode column with embedded strong basic ion-pairing groups, using an isocratic analytical method with a simple mobile phase of water, acetonitrile (MeCN, ACN), and sulphuric acid (H2SO4) buffer. UV detection at 240 nm.
Column | Primesep SB, 3.2×100 mm, 5 µm, 100A |
Mobile Phase | MeCN/H2O – 20/80% |
Buffer | H2SO4 – 0.2% |
Flow Rate | 1.0 ml/min |
Detection | 240 nm |
Class of Compounds | Drug, Hydrophobic |
Analyzing Compounds | Escitalopram |
Application Column
Primesep SB
The Primesep family of mixed-mode columns offers a wide variety of stationary phases, boasting unprecedented selectivity in the separation of a broad array of chemical compounds across multiple applications. Corresponding Primesep guard columns, available with all stationary phases, do not require holders. SIELC provides a method development service available to all customers. Inquire about our specially-tailored custom LC-phases for specific separations.
Select options