Pyridine is an aromatic heterocyclic organic compound with the chemical formula C₅H₅N. It is structurally related to benzene, with one methine group (=CH−) replaced by a nitrogen atom.
Pyridine can be retained, separated and analyzed using a Primesep 100 mixed-mode stationary phase column. The analysis employs a gradient method with a simple mobile phase comprising water, acetonitrile (MeCN), and sulfuric acid as a buffer. This method allows for detection using UV 250 nm.
You can find detailed UV spectra of Pyridine and information about its various lambda maxima by visiting the following link.
Column | Primesep 100, 4.6 x 150 mm, 5 µm, 100 A |
Mobile Phase | MeCN – 70% |
Buffer | H2SO4 – 0.05% |
Flow Rate | 1.0 ml/min |
Detection | UV 250 nm |
Limit of Detection | 5 ppb |
Class of Compounds | Heterocyclic aromatic |
Analyzing Compounds | Pyridine |
Application Column
Primesep 100
Column Diameter: 4.6 mm
Column Length: 150 mm
Particle Size: 5 µm
Pore Size: 100 A