HPLC Method for Analysis of 1-Ethylpyridinium Bromide on Newcrom AH Column

Separation type: Liquid Chromatography Mixed-mode

HPLC Method for Analysis of 1-Ethylpyridinium bromide on Newcrom AH by SIELC Technologies

HPLC Method for Analysis of 1-Ethylpyridinium Bromide on Newcrom AH Column
HPLC Method for Analysis of 1-Ethylpyridinium bromide on Newcrom AH by SIELC Technologies

1-Ethylpyridinium bromide is a type of ionic liquid. An ionic liquid is a salt where the ions are poorly coordinated, resulting in these solvents being liquid below 100°C, or even at room temperature (ambient temperature). At least one ion will have a delocalized charge and one component will be organic, which prevents the formation of a stable crystal lattice.

In 1-ethylpyridinium bromide, the cation is an ethylpyridinium ion, a pyridine ring (an aromatic six-membered ring with five carbon atoms and one nitrogen atom) with an ethyl group (a two-carbon alkyl chain) attached. The anion is chloride (Cl-).

Ionic liquids, including 1-ethylpyridinium chloride, have a wide range of applications, such as in electrochemistry, catalysis, and as solvents for a variety of chemical reactions. Their properties, including melting point, viscosity, and solvation potential, can be modified by selecting different cations and anions, which gives them great versatility

The 1-Ethylpyridinium bromide can be retained, analyzed, on a Newcrom AH column using an isocratic analytical method with a simple mobile phase of water, Acetonitrile (MeCN), and an Ammonium formate (AmFm) ionic modifier. This analysis method can be detected with an Evaporative Light Scattering Detector (ELSD) or any other evaporative detection method (CAD, ESI-MS).

LOD was determined for this combination of instrument, method, and analyte, and it can vary from one laboratory to another even when the same general type of analysis is being performed.

High Performance Liquid Chromatography (HPLC) Method for Analysis of  1-Ethylpyridinium bromide

Condition

ColumnNewcrom AH, 4.6 x 100 mm, 5 µm, 100 A
Mobile PhaseMeCN/H2O – 50/50%
BufferAmFm pH 3.0- 10 mM
Flow Rate1.0 ml/min
DetectionELSD, the nebulizer and evaporator temperatures 50 °C, with a gas flow rate of 1.6 Standard Liters per Minute (SLM)
(MS- compatible mobile phase)
Peak Retention Time4.92 min
Sample concentration0.3 mg/ml
Injection volume1 µl
LOD UV- 30 ppb, ELSD – 70 ppb

Description

Class of CompoundsPyridinium salts
Analyzing Compounds1-Ethylpyridinium bromide

Application Column

Newcrom AH

Column Diameter: 4.6 mm
Column Length: 100 mm
Particle Size: 5 µm
Pore Size: 100 A

Add to cart
Application Analytes:
1-Ethylpyridinium bromide

Application Detection:
ELSD Detection
SIELC Technologies usually develops more than one method for each compound. Therefore, this particular method may not be the best available method from our portfolio for your specific application. Before you decide to implement this method in your research, please send us an email to research@sielc.com so we can ensure you get optimal results for your compound/s of interest.